38 MS-70 / MX-50 / MF-50 / ML-50
Select an Analysis Mode
5. Press the SELECT key to blink a symbol, it will
blink. Select an analysis mode using the ↓ or ↑ key.
If standard mode is used, select 5td .
If automatic mode is used, select U-a .
If manual mode is used, select U-m .
Set ACCURACY for the Standard Mode
6. Select ACCURACY with the SELECT key.
Select MID. with the ↓ or ↑ key.
When ACCURACY is selected, HI, MID. or LO blinks.
Set the Heating Pattern
7. Select drying temperature 1 with the SELECT
key. Select step drying of the heating
pattern with the PROGRAM key.
Set Drying Temperature 1
8. Set 120°C to drying temperature 1 with
the ↓ or ↑ key.
Set the Time
9. Select time with the SELECT key.
Set 5.0 minutes with the ↓ or ↑ key.
Set Drying Temperature 2
10. Select drying Temperature 2 with
the SELECT key.
Set 160°C with the ↓ or ↑ key.
Set the Preset Termination Value for Automatic Mode
11. Select a preset termination value with the
SELECT key. Select 0.05 [%/min] with
the ↓ or ↑ key.
Set Measurement Unit
12. Select measurement unit with the SELECT key.
Select moisture content (based on a wet
sample ) with the
↓ or ↑ key.
ACCURAC
MID.
Standard
mode
"0."0"5
%/min
Automatic
mode
"U"-"a
0.000g
0"%
"M"O"I"S"T
"/"W
"5."0 min
2
"1"6"0°C
"1"0"5°C
1
"1"2"0°C
Standard mode
Manual mode
Automatic mode
Manual mode
ACCURACY
Preset termination value
Analysis
Measurement unit
Drying temperature 2
Drying temperature 1
Step drying
Time